| |
All
operating room staff should be aware of the risks of fire
or explosion as a result of the use of anaesthetic vapours.
It is important to distinguish between gas mixtures that
can burn and those that are explosive. Explosions are much
more dangerous to both staff and patients. Of the inhalational
anaesthetics mentioned in this book, only ether is flammable
or explosive in clinical concentrations:
| :: |
Mixtures
of ether and air in the concentrations used for anaesthesia
are flammable |
| :: |
Mixtures
of ether and air (whatever the concentration) are not
explosive |
| :: |
Mixtures
of ether with oxygen or nitrous oxide are explosive |
| :: |
Other
substances used in the operating room, such as alcoholic
skin preparations, also present a risk of fire or even
explosion in the presence of high concentrations of oxygen. |
There is no site within the draw-over apparatus or breathing
system where a fire or explosion could start. The point of
risk is therefore the place where the patient’s expired
gas enters the room – or at any point where anaesthetic
gas accidentally leaks into the room from the apparatus. If
you are using 3–5% ether as an anaesthetic in combination
with a muscle relaxant, it is likely that the ether concentration
in the patient’s expired gas will be less than the
lowest flammable concentration (2%).
If ether is used on a compressed
gas machine (Boyle’s
machine), the gases are always explosive.
When flammable gases are in use, the most likely sources of
combustion in the operating theatre are the surgical diathermy
machine and other electrical apparatus. Static electricity
is unlikely to start a fire, but may trigger an explosion if
an oxygen-rich gas mixture is present.
To minimize the risk of explosion, never allow the simultaneous
use of diathermy on a patient anaesthetized with ether. If
one of these techniques must be used for the benefit of the
patient, the other must not be allowed.
If possible, your operating room and equipment should be of
the antistatic type. This is important in a dry climate, but
less so in a humid one where a natural coating of moisture
on objects prevents the buildup of static.
Electrical sockets and switches should either be spark-proof
or be situated at least 1 metre above floor level. The patient’s
expired gases should be carried away from the expiratory valve
down wide-bore tubing at least to the floor (ether is heavier
than air) or out of the operating room. Make sure that no-one
stands on the hose and that there is nothing that could trigger
combustion near the end of the tubing. If you use oxygen enrichment
during induction, but not surgery, the patient’s expired
gas will cease to be explosive within 3 minutes of stopping
the addition of oxygen.
No potential cause of combustion or source of sparking should
be allowed within 30 cm of any expiratory valve through which
a potentially flammable or explosive mixture is escaping.

|
|
|